Late partial dissociation of a humeral head morse taper associated with periprosthetic proximal humeral fracture a case report

Alan H. Lee, Matthew Dilisio, Jon J P Warner

Research output: Contribution to journalArticle

Abstract

A fifty-two-year-old man who had undergone total shoulder arthroplasty ten years previously presented following a mechanical fall onto his operatively treated shoulder, resulting in a periprosthetic humeral fracture and partial dissociation of the humeral head from the stem. Conclusion: Shoulder surgeons should consider late disengagement of the Morse taper as a potential cause of surgical failure. Partial late dissociation of the humeral head from the stem after shoulder arthroplasty is possible and can be associated with a fracture. The shoulder surgeon should be aware of this potential complication when evaluating periprosthetic proximal humeral fractures, emphasizing the importance of attention to detail when evaluating patients.

Original languageEnglish
Article numbere34
JournalJBJS Case Connector
Volume6
Issue number2
DOIs
StatePublished - 2016

Fingerprint

Shoulder Fractures
Humeral Head
Arthroplasty
Periprosthetic Fractures
Humeral Fractures

All Science Journal Classification (ASJC) codes

  • Surgery
  • Orthopedics and Sports Medicine

Cite this

Late partial dissociation of a humeral head morse taper associated with periprosthetic proximal humeral fracture a case report. / Lee, Alan H.; Dilisio, Matthew; Warner, Jon J P.

In: JBJS Case Connector, Vol. 6, No. 2, e34, 2016.

Research output: Contribution to journalArticle

@article{c684f1ad0df04a1fad1d012d98472aef,
title = "Late partial dissociation of a humeral head morse taper associated with periprosthetic proximal humeral fracture a case report",
abstract = "A fifty-two-year-old man who had undergone total shoulder arthroplasty ten years previously presented following a mechanical fall onto his operatively treated shoulder, resulting in a periprosthetic humeral fracture and partial dissociation of the humeral head from the stem. Conclusion: Shoulder surgeons should consider late disengagement of the Morse taper as a potential cause of surgical failure. Partial late dissociation of the humeral head from the stem after shoulder arthroplasty is possible and can be associated with a fracture. The shoulder surgeon should be aware of this potential complication when evaluating periprosthetic proximal humeral fractures, emphasizing the importance of attention to detail when evaluating patients.",
author = "Lee, {Alan H.} and Matthew Dilisio and Warner, {Jon J P}",
year = "2016",
doi = "10.2106/JBJS.CC.15.00111",
language = "English",
volume = "6",
journal = "JBJS Case Connector",
issn = "2160-3251",
publisher = "Lippincott Williams and Wilkins",
number = "2",

}

TY - JOUR

T1 - Late partial dissociation of a humeral head morse taper associated with periprosthetic proximal humeral fracture a case report

AU - Lee, Alan H.

AU - Dilisio, Matthew

AU - Warner, Jon J P

PY - 2016

Y1 - 2016

N2 - A fifty-two-year-old man who had undergone total shoulder arthroplasty ten years previously presented following a mechanical fall onto his operatively treated shoulder, resulting in a periprosthetic humeral fracture and partial dissociation of the humeral head from the stem. Conclusion: Shoulder surgeons should consider late disengagement of the Morse taper as a potential cause of surgical failure. Partial late dissociation of the humeral head from the stem after shoulder arthroplasty is possible and can be associated with a fracture. The shoulder surgeon should be aware of this potential complication when evaluating periprosthetic proximal humeral fractures, emphasizing the importance of attention to detail when evaluating patients.

AB - A fifty-two-year-old man who had undergone total shoulder arthroplasty ten years previously presented following a mechanical fall onto his operatively treated shoulder, resulting in a periprosthetic humeral fracture and partial dissociation of the humeral head from the stem. Conclusion: Shoulder surgeons should consider late disengagement of the Morse taper as a potential cause of surgical failure. Partial late dissociation of the humeral head from the stem after shoulder arthroplasty is possible and can be associated with a fracture. The shoulder surgeon should be aware of this potential complication when evaluating periprosthetic proximal humeral fractures, emphasizing the importance of attention to detail when evaluating patients.

UR - http://www.scopus.com/inward/record.url?scp=84977263765&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84977263765&partnerID=8YFLogxK

U2 - 10.2106/JBJS.CC.15.00111

DO - 10.2106/JBJS.CC.15.00111

M3 - Article

VL - 6

JO - JBJS Case Connector

JF - JBJS Case Connector

SN - 2160-3251

IS - 2

M1 - e34

ER -