Abstract
An isocratic technique was developed for the analysis of leuprolide acetate by high performance liquid chromatography (HPLC) using 220 nm UV detection, C18 MICROSORB-MV™ column (4.6 mm×15 cm), mobile phase (0.03 M, 77% dibasic ammonium phosphate buffer: 23% acetonitrile), and 2.0 mL/min flow rate. The coefficient of variation (C.V.) for precision and proportionality assays was lower than 2% for all concentrations studied. The detection limit and tailing factor for leuprolide acetate were 100 ng/mL (signal-to-noise ratio of 3:1) and 1.16-1.26, respectively.
Original language | English (US) |
---|---|
Pages (from-to) | 3023-3031 |
Number of pages | 9 |
Journal | Journal of Liquid Chromatography and Related Technologies |
Volume | 23 |
Issue number | 19 |
DOIs | |
State | Published - Dec 1 2000 |
Externally published | Yes |
All Science Journal Classification (ASJC) codes
- Analytical Chemistry
- Biochemistry
- Pharmaceutical Science
- Clinical Biochemistry